BCA1000 Automatic Biochemical Analyzer

Description:

Overall Performance

Equipment type Fully automatic discrete, STAT priority

Analysis rate Constant speed 240T/H

Test principle Colorimetry, turbidimetry

Analysis method End-point, 2 kinetics, fixed-time, etc. support single/double wavelength, linear

and non-linear calibration

Description

BCA1000 Automatic Biochemical Analyzer

BCA1000 Automatic Biochemical Analyzer

Sample System

Tray 60 sample positions, random access of all samples

Volume 2-50ul, 0.1ul increment

probe Liquid level detection, the remaining detection, collision protection etc.

Sample Containers Sample cups, Micro tube, Original blood-collection tube, etc.

Reagent System

Tray 45 reagent positions

Volume R1:25-300ul, R2:10150ul, 0.1ul increment

Probe Liquid level detection, the remaining detection, collision protection etc.

Reagent Refrigeration 415℃, independent power supply, 24-hour non-stop refrigeration

Reaction System

Reaction Tray Rotating tray, 120 cuvettes

Cuvette Reusable high quality cuvette, Optical diameter 6mm

Reaction Volume 180-500ul

Incubation system Bi-directional liquid-air device

Temperature 37℃±0.1℃ with real-time display

Optical System

Light source 12V35W halogen lamps, over 2000 hours

Photoelectron road after spectrophotometry

Wavelength 340,405,450,510,546,578,630,700nm

Absorbance range 0-3.500Abs

Resolution 0.0001Abs

Operating System

PC operating system Windows XP or above

Analysis control software English version graphical operating software

Report printing Report formats support the user-defined mode, QC and state information etc.

System connection RS-232

Working Conditions

Power supply AC 220V±22V, 50Hz±1Hz, power 1.5KVA

Ambient Temperature 15-30 ℃

Humidity 35-80%

Water consumption 6L/Hour

Dimension Bench top: 850mm*650mm*690mm

Weight 80KG