Respiratory Ventilator Medical PA 700B
R242,500.00 Ex VAT
Advanced Model Exclusive Features
Equipped with professional medical grade air compressor tailored for long clinical times and low noise requirements
This product is strictly by order only. Delivery is by air, from Dubai. Price does not include shipping Please allow 10 to 15 working days delivery. ( Delays is highly possible )
1 in stock
Description
Respiratory Ventilator Medical PA 700B
Ventilator with a medical air compressor(PA-700B)
- The Ventilator is a universal multi-mode Ventilator, a multi-functional Ventilator, which can be used for clinical breath first aid and respiratory therapy.
- With the control, auxiliary/control, intermittent instruction, manual and other breathing patterns, as well as breathing is not positive pressure, sigh and other respiratory function is the hospital’s essential breathing equipment.
- Large-screen LCD display, 10.4 inch TFT color LCD screen. Dual flow sensor to ensure that the breath and inhale more secure, accurate.
- Power-off-delay function.


| Display | 10.4” LCD Display |
| Mode: | a pneumatically powered and electrically controlled system |
| Trolley: | Fitted with 4 nos anti-static rubber castors; two of which
are lockable for braking and easy maneuverability with foot-operated brake provisions |
| Technical Specifications: | |
| Gas requirement: | Medical oxygen and nitrous oxide with a pressure ranging from 280~600kpa |
| safety valve | <12.5 kPa |
| Respiratory Rate | 1 ~ 99 bpm |
| Oxygen Concentration | 21% ~ 100% |
| Maximum Minute ventilation | ≥18L / min |
| modes of ventilation | A/C, IPPV, SIPPV, IMV, SIMV, PEET. MANUAL, SIGH, SPONT |
| Ventilator compliance | ≤30 Ml/kPa |
| PEEP: | 0 ~ 1.0 kPa |
| Ptr: | -10 ~ 10 hPa |
| IMV frequency: | 1~20 bpm |
| I/E Ratio: | 4:1~1:4, Has the inverse ratio ventilation |
| Tidal Volume | Setting Range: 50~1500ML
Display Range: 0~2000ml ; resolution ratio: 10ml |
| Inspiratory Plateau: | 0~1s |
| Maximum Safety Pressure | ≤ 6.0kPa |
| Sigh: | one deep breath per 80 controlled respirations,
the inspiration time is 1.5 times of the setting point |
| Pressure limit range: | 1.0~ 6.0kPa |
| Airway pressure alarm: Audible and visual and
with yellow and red color indicating |
Lower: 0.1 kPa ~ 6.0 kPa; Upper: 0 ~ 5.0 kPa |
| ±0.2 kPa | |
| Ventilation Volume alarm: | ≥ 18L/ MIN |
| Power Supply Alarm | Ac/dc power supply are after failing to send out alarm immediately Alarm Time: keep >120s |
| Operating conditions | |
| Ambient temperature: | 10 ~ 40oC |
| Relative humidity: | no higher than 80 % |
| Atmospheric pressure: | 860 hPa ~ 1060 hPa |
| Power requirement: | 100-120 Vac, 50/60 Hz; |
| Attention: the AC power supply used for the anesthesia
the machine must be well-grounded. |
|
| Storage | |
| Ambient temperature: | -15oC ~ +50oC |
| Relative humidity: | no higher than 95 % |
| Atmospheric pressure: | 86 kPa ~ 106 kPa. |
| It should be stored in a room without corrosive gas and well ventilated | |
| Package | |
| packaging box | comply with the requirement of GB/T 15464 |
| Between the packaging box and product, soft material
with appropriate thickness provided to prevent loosening and mutual friction during transportation |
|
| Moisture protection and rain protection to ensure that
product is protected from natural damage. |
|
| Safety & Alarm | |
| Gas source pressure | Immediate alarm; “high level alarm” indicator flickers raising an audible alarm. |
| Airway Pressure Alarm | The airway pressure exceeds 15hpa. The maximum delay is no longer than 17s. The “high-level alarm” indicator flickers raising an audible alarm. |
| Tidal Volume Alarm | Lower than 50ml, The alarm will be raised after one breathing cycle. |
| Ventilation Volume Alarm | Lower: 0~12L/Min; Up: 18L/ Min |
| Battery Voltage Alarm | It alarms when the battery voltage is lower than 10.5V. |
| Power On Self Test | Self-test before working, assure it’s working normally. |