Syringe Pump KL-605T

R11,621.80 Ex VAT

Specifications

Model KL-605T
Syringe Size 5, 10, 20, 30, 50/60 ml
Applicable Syringe Compatible with the syringe of any standard
VTBI 0.1-1200 ml (in 0.1, 1, 10 ml increments)
Flow Rate Syringe 5 ml: 0.1-100 ml/h (in 0.01, 0.1, 1, 10 ml/h increments)Syringe 10 ml: 0.1-300 ml/hSyringe 20 ml: 0.1-600 ml/h

Syringe 30 ml: 0.1-800 ml/h

Syringe 50/60 ml: 0.1-1200 ml/h

Bolus Rate 5 ml: 0.1-100 ml/h (in 0.01, 0.1, 1, 10 ml/h increments)10 ml: 0.1-300 ml/h20 ml: 0.1-600 ml/h

30 ml: 0.1-800 ml/h

50/60 ml: 0.1-1200 ml/h

Anti-Bolus Automatic
Accuracy ±2% (mechanical accuracy≤1%)
Infusion Mode Memory modeFlow rate: ml/min, ml/hTime-based

Bodyweight

KVO Rate 0.1-1 ml/h (in 0.01 ml/h increments)
Alarms Occlusion, near empty, end program, low battery, end battery,AC power off, motor malfunction, system malfunction, reminder alarm,pressure sensor error, syringe installation error, syringe drop off
Additional Features Real-time infused volume, automatic power switching,automatic syringe identification, mute key, purge, bolus, anti-bolus,system memory, history log
Drug Library Available
Occlusion Sensitivity High, medium, low
History Log 50000 events
Power Supply, AC 110/230 V, 50/60 Hz, 20 VA
Battery 14.8 V, rechargeable
Battery Life 8 hours at 5 ml/h
Working Temperature 5-40℃
Relative Humidity 20-90%
Atmospheric Pressure 700-1060 hpa
Size 245*120*115 mm
Weight 2.5 kg
Safety Classification Class Ⅱ, type BF

1 in stock

Description

Syringe Pump KL-605T

 

Description

Syringe Pump KL-605T

Intravenous infusion pump

Should be compact weighing 2.5 Kg color TFT display Syringe sizes 5,10,20,30,50/60 ml Should be compatible with syringe of any brand standard Flow rate range Syringe 5 ml 0.1-100 ml/h in 0.01, 0.1, 1, 10 ml/h increments Syringe 10 ml 0.1-300 ml/hSyringe 20 ml 0.1-600 ml/h Syringe 30 ml 0.1-800 ml/h Syringe 50/60 ml 0.1-1200 ml/h Should have Bolus rate5 ml 0.1-100 ml/h in 0.01, 0.1, 1, 10 ml/h increments10 ml 0.1-300 ml/h20 ml 0.1-600 ml/h30 ml 0.1-800 ml/h50/60 ml 0.1-1200 ml/h Anti-Bolus Mechanism Flow rate setting ml/Min, ml/h Should have infusion modes- Flow based, Time based, body weight mode Should have Body weight rate calculation in g/kg/min, g/kg/h, mg/kg/min, mg/kg/h Should have KVO rate 0.1-1 ml/h in 0.01 ml/h increments Should have alarms for Occlusion, Near empty, End programme, low battery, A/C power off, motor malfunction, System malfunction, Pressure sensor error, Syringe drop off.